Before Using this Product
28
PANOURA 18S Operation Manual
1.3.4 Names and functions of hand-switch box indicators and switches
2
3
4
1
5
Name
Function
1
X-ray LED
Keeps on during X-ray radiation with a continual acoustic alarm.
2
Ready LED
Blinks while the equipment (e.g. the rotating arm) is operating; goes on when
it has returned to its reset point in each radiation mode; goes off when it
leaves the reset point. Blinks while the arm is moving to the positioning beam
radiation point; goes off when it has reached the point.
3
Cooling LED
Keeps on during cooling time after X-ray radiation; goes off when the cooling
time has expired.
4
Power LED
Keep on while power is being supplied.
5
X-ray radiation switch
Pressed and held down in the Ready for radiation state (Ready LED on) to
irradiate X-ray for a specified time period.
*In case of an error, X-ray Ready LED/Cooling LED blinks. When the problem is fixed, The LED goes off.
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......