1.3.6 Positioning beams
Positioning for Panoramic Image Acquisition
Pressing any of the positioning beam switches activates its associated positioning beam (at
the same time, the arm unit moves to the positioning point). These positioning beams can be
used to make the patient’s head immovable at the position appropriate for image acquisition.
1
2
3
Name
Function
1
Median beam
Used to adjust any horizontally inclined patient's face by aligning to
the patient's midsagittal.
2
Oculoauricular beam
Used to adjust any anteroposteriorly inclined patient's face by align-
ing to the patient's reference face plane (Frankfurt or Camper plane).
3
Front-tooth tomographic beam
Used to adjust the patient's front tooth row and the X-ray radiation
point by aligning the beam to the tops of front tooth roots (around the
middle of the mandibular third tooth).
Warning
• Don’t gaze directly at the positioning beams, though their power outputs (Class 2) are
enough low to avoid adverse effects on eyes by aversive behavior, e.g. winking.
Caution
• Take care not to irradiate the laser positioning beams into the patient’s eyes. Ask the
patients not to gaze at these beams.
Suppl.
: The positioning beams automatically go off about 60 seconds after activation. If
the positioning beams have gone off before positioning is successfully finished,
retry pressing the beam switches to turn the positioning beams on?
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......