![background image](http://html1.mh-extra.com/html/yoshida/panoura-18s/panoura-18s_operation-manual_3502917111.webp)
Recovering from failures
109
PANOURA 18S Operation Manual
6.1.2 Problems associated error messages displayed on the screen
If an error message appears on the control panel, follow the corrective measures described
in "Error message code list" to fix.
Error message code list
Error code
Description
Action
Er08
The X-ray radiation switch has been released during image
acquisition.
Press the Reset switch on the con-
trol panel to recover from the error.
Retry.
Er09
The rotating arm has rotated faster than the given speed.
If this error persists, contact your
dealer.
Er10
The rotating arm has rotated slower than the given speed.
Er27
FLASH memory writing error
Er28
X-ray cannot be irradiated at the specified strength
Immediately contact your dealer.
Er29
X-ray cannot be irradiated at the specified dose.
Er30
X-ray head temperature has exceeded the given value.
If this error persists, contact your
dealer.
Er31
Failure in the X-ray head
Immediately contact your dealer.
Er34
Failure in arm sensor AR
Er39
Failure in arm sensor AL
Er44
Two or more arm driving sensors including the arm sensor
have been activated at the same time.
Er46
Communication error between CPU circuit board and touch
panel
If this error persists, contact your
dealer.
Er50
At main power-on, the X-ray radiation switch has been
pressed in the hand switch box.
Press the Reset switch on the
control panel to recover from the
error. If this error persists, contact
your dealer.
Er54
The arm unit movement does not finish within the given
time period.
Er56
The back-and-forth driving motor movement does not finish
within the given time period.
Er57
Failure in back-and-force driving sensor. A wrong combina-
tions of arm driving sensors have been activated at the
same time.
Immediately contact your dealer.
Er59
Motor control error. Image acquisition control has abnor-
mally terminated.
If this error persists, contact dealer.
Er60
Arm unit driving control error. Arm unit driving motor control
has abnormally terminated.
Er62
Back-end-forth driving motor error. Back-and-forth driving
motor control has abnormally terminated.
Er67
A major difference has observed between the arm unit
driving pulse and the value detected by the arm position
detector.
Press the Reset switch on the
control panel to recover from the
error. If this error persists, contact
your dealer.
Er68
The value detected by the arm position detector has ex-
ceeded the normal range of operation due to running idle.
Er69
The timing between the back-and-forth driving motor pulse
and the toggling of back-and-forth driving sensor ON and
OFF has exceeded the normal range.
Er71
X-ray head control error
Er72
The type of the X-ray head has not correctly specified.
Immediately contact your dealer.
Er73
Communication error between CPU circuit board and hand
switch box
Press the Reset switch on the
control panel to recover from the
error. If this error persists, contact
your dealer.
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......