Recovering from failures
108
PANOURA 18S Operation Manual
Symptom
Possible cause
Action
Abnormal noise sounds
during X-ray radiation.
The X ray tube has been shorted out or a
gas has been generated in it.
Contact your dealer.
The high-voltage transformer coil has
been shorted out.
No reaction to opera-
tion from the PC
The mouse keyboard has not been cor-
rectly connected to the PC.
Connect the mouse keyboard connector
to the PC.
No power is supplied to
the PC.
The power cable has not been connected
to the PC.
Plug in the power cable firmly into both
the PC and the outlet.
None appears on the
monitor screen.
The cable has not been correctly con-
nected between the monitor and the PC.
Connect correctly the cable to the PC.
The power cable has not been connected
to the monitor.
Plug in the power cable firmly into both
the monitor the outlet.
The PC has been activated but the power
switch of the monitor has not been turned
on.
Check that the power button of the
monitor has been pressed and power is
supplied (ON).
"Check that power is
supplied to the main
body or the cable has
been correctly connect-
ed to the PC" message
has appeared.
The PC connection cable has not been
correctly connected to the PC.
Connect correctly the cable to the PC.
The PC communication port has not been
configured.
Configure the communication port (see
"Configuring a communication port(s)."
P.101 ).
No power is supplied to the unit.
Turn the main switch on (a | symbol will
appear).
Failed unit
Contact your dealer.
Troubleshooting of Image acquisition
(For a Cephalo-attached equipment only)
Symptom
Possible cause
Action
The equipment does
not switch to Cephalo
image acquisition
mode.
Arm does not move to Cephalo image
acquisition position.
Contact your dealer if it happens fre-
quently.
CCD sensor unit does not move to
Cephalo image acquisition position,
Some parts of the
Cephalo image is
missing (radiation field
is off).
Slit position or Cephalo sensor position is
not adjusted properly.
Contact your dealer.
Cephalo image is
blurred.
Patient may have moved during the image
acquisition.
Fix the patient properly and advise him
not to move during image acquisition.
Right and left ear
rods position for the
Cephalo lateral image
(LA) is off from the
original position.
Something may have hit the ear rod.
Contact your dealer.
Содержание PANOURA 18S
Страница 2: ......
Страница 23: ...Before Using this Product 21 PANOURA 18S Operation Manual...
Страница 24: ...Before Using this Product 22 PANOURA 18S Operation Manual...
Страница 44: ...MEMO...
Страница 108: ...MEMO...
Страница 118: ...MEMO...
Страница 120: ...MEMO...
Страница 122: ...MEMO...
Страница 124: ...MEMO...
Страница 129: ...Main specification 127 PANOURA 18S Operation Manual 11 3 2 Standing position wall mount short type Panorama type Unit mm...
Страница 130: ...Main specification 128 PANOURA 18S Operation Manual 11 3 3 Standing position wall mount long type Panorama type Unit mm...
Страница 137: ...Main specification 135 PANOURA 18S Operation Manual 11 3 10 Standing position wall mount Cephalo attached type Unit mm...
Страница 150: ...MEMO...
Страница 154: ...MEMO...
Страница 155: ......