16-10
IM 760401-01E
Item
Specifications
Unit symbols
m, k, M, V, A, W, VA, var, deg, Hz, h
±
, TIME, %
Number of displayed digits Select 5 or 4 digits
Display update rate
During normal measurement
Select 0.1 s, 0.25 s, 0.5 s, 1 s, 2 s, or 5 s.
During harmonic measurement
Select 0.25 s, 0.5 s, 1 s, 2 s, or 5 s.
Response time
Approx. (display update rate + 100 ms) (The time it takes to reach the accuracy of the final value
when the displayed value changed from 0 to 100% or 100 to 0% of the rated range)
Auto range monitor
The indicator illuminates when the input signal meets the conditions for auto range switching.
Overrange display
Overrange “- - oL-” is displayed for the following conditions.
During normal measurement
When the measured value exceeds 140% of the rated range
During harmonic measurement
•
For 600-V range or 20-A range (300 V range or 10 A range if the crest factor is set to 6)
When the measured value exceeds 140% of the rated range
•
For other ranges
When the measured value exceeds 200% of the rated range
Hold
Holds the displayed value.
Single update
Updates the displayed value once each time the TRIG key is pressed during Hold.
MAX hold
Holds the maximum displayed value of V, A, W, VA, var, voltage peak, and current peak.
Internal memory
Item
Specifications
Measured data
Number of stored data points
Product
Normal
Harmonic
measurement
measurement
WT210 (single-phase model, 760401)
600 blocks
30 blocks
WT230 (three-phase, three-wire model, 760502)
300 blocks
30 blocks
WT230 (three-phase, four-wire model, 760503)
200 blocks
30 blocks
Store interval
Display update rate or in the range of 1 s to 99 hrs 59 min 59 s.
Recall interval
Display update rate or in the range of 1 s to 99 hr 59 min 59 s.
Setup information
Stores/Recalls four patterns of setup information.
16.3 Functions
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...