
7-7
IM 760401-01E
Harmonic Measurement Function (Option)
7
7.4
Setting the Harmonic Order and Displaying the
Measured Harmonic Value
Keys
CHECK RANGE
MODE
UPDATE
SCALING
AVG
LINE
FREQ
STORE
RECALL
HARMONICS KEY LOCK
FILTER
A
B
C
VOLTAGE
VOLTAGE
VOLTAGE
AUTO
MODE
CURRENT
AUTO
RANGE
MAX HOLD
TRIG
CAL
INTEGRATOR
INTEG SET
SHIFT
KEY LOCK
1P 3W
3P 3W
3P 4W
3V 3A
OUTPUT
HARMONICS
MEMORY
REMOTE
MEAN
CURRENT
RMS
DC
FUNCTION
ELEMENT
MAX HOLD
HOLD
ENTER
RESET
WIRING
START
STOP
LOCAL
SETUP
1
m
V
k
A
V
TIME
ar
deg
h
h
VA
M W
m
V
PF
k
A
M W
m
V
Hz
k
A
M W
2
3
FUNCTION
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
%
The explanation given in this section uses WT230 as an example. For the differences between
the WT210 and the WT230, see section 2.2, “Operation Keys and Functions/Element Display.”
Procedure
The following operations assume that the harmonic measurement function is turned ON.
Setting the Harmonics Order
1.
Light up the display function indicator of display A.
Display
A
Light up display function
V
A
W
FUNCTION
FUNCTION
FUNCTION
FUNCTION
2.
Set the harmonics order.
(Display A)
Displaying the Measured Harmonic Values
Displays each measured value after having set the display function of either display B
or C.
FUNCTION
V
A
W
PF
FUNCTION
FUNCTION
FUNCTION
FUNCTION
FUNCTION
V% A%
W%
Vdeg
A%
V%
Adeg
FUNCTION
FUNCTION
FUNCTION
FUNCTION
FUNCTION
Display
B
FUNCTION
V
A
W
V Hz
FUNCTION
FUNCTION
FUNCTION
FUNCTION
A Hz
Display
C
Explanation
Setting the Order of Harmonics
The maximum order for which analysis results can be displayed varies depending on
the frequency of the fundamental.
Example
• When the fundamental frequency is 50 Hz, up to 50 orders can be displayed;
• When the fundamental frequency is 400 Hz, up to 30 orders can be displayed.
When an order is set exceeding the maximum order, display B will change to the dot
display. Refer to Chapter 16 for more details on upper limits of analysis orders.
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...