
6-11
IM 760401-01E
Integration
6
Integration Computation when the Measured Value Exceeds Measurement Limits
Integration Computation When the Measured Value Exceeds the Measurement Limit
When the instantaneous voltage or current of the sampled data exceeds approx. 333%
(approximately 666% if the crest factor is set to 6) of the measurement range, the value
is considered to be the value corresponding to approx. 333% (approximately 666% if the
crest factor is set to 6) for the purpose of integration.
Integration When Current Input Is Small
When the measurement mode is RMS/VOLTAGE MEAN and the current input drops
below 0.5% (less than or equal to 1% if the crest factor is set to 6) of the rated range, the
ampere hour value is integrated as zero (0).
Valid Frequency Range for Integration
The sample rate is approximately 50 kHz. The voltage/current signal frequencies that
are valid for the integration are as follows.
Integrated Item
Valid Frequency Range for Integration
Active power
DC to 25 kHz
Current
When the measurement mode is RMS
DC, lower limit frequency determined by the
display update rate to 25 kHz
When the measurement mode is VOLTAGE MEAN
DC, lower limit frequency determined by the
display update rate to 25 kHz
When the measurement mode is DC
DC to 25 kHz
6.4 Precautions Regarding Use of Integrator Function
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...