10-2
IM 760401-01E
GP-IB Interface Specifications
• Electrical & mechanical specifications: conforms to IEEE st’d 488-1978
• Functional specifications: refer to the table below
• Protocol: Varies depending on the mode used. See the previous page.
• Code: ISO (ASCII) code
• Address setting: 0 to 30 listener and talker addresses, or talk-only can be selected
using the front panel keys.
• Remote mode clear: remote mode can be cleared by pressing the LOCAL key on the
front panel. However, this is not possible when Local Lockout has been set by the
controller.
Function
Subset name
Description
source handshake
SH1
full source handshake capability
acceptor handshake
AH1
full acceptor handshake capability
talker
T5
basic talker capability, serial polling, nontalker on MLA
(My Listen Address), talk-only capability
listener
L4
Basic listener capability, nonlistener to MTA (My Talk
Address), no listen-only capability
service request
SR1
full service request capability
remote local
RL1
full remote/local capability
parallel poll
PR0
no parallel polling capability
device clear
DC1
full device clear capability
device trigger
DT1
full device trigger capability
controller
C0
no controller function
WARNING
The connectors used in this function have protective covers. When the covers
are removed or when using connectors, the voltage ratings across the
measuring input and the ground become as follows:
Voltage between CURRENT,
±
(VOLTAGE and CURRENT side) input
terminals and ground
400 Vrms max.
Voltage between VOLTAGE input terminal and ground
600 Vrms max.
Put the protective cover on the connector when this function is not used.
10.1 GP-IB Interface Functions and Specifications
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...