15-9
IM 760401-01E
Maintenance and Troubleshooting
15
Preparation
Set the frequency of the AC voltage standard and of each channel of the synchronizer
to 60 Hz. Then, while not exceeding the maximum values of the external
synchronization inputs of the voltage and current standard, rise the output level of the
synchronizer until the standards are synchronized.
Set the crest factor to 3.
Calibrating
• Calibration of voltage and current
1.
Set the voltage range and current range of the WT210/WT230 to the range to be
calibrated.
2.
Set the output voltage of the AC voltage standard so that the voltage reading on
the standard power meter matches the rated voltage range of the WT210/
WT230.
3.
Set the output current of the AC current standard so that the current reading on
the standard power meter matches the rated current range of the WT210/
WT230.
The voltage and current values displayed on the WT210/WT230 at this point are the
calibration values.
• Calibration of power when the power factor = 1
4.
Shift the phase of the current signal channel of the phase shifter so that the
power indicated on the standard power meter shows the rated value.
The power value displayed on the WT210/WT230 at this point is the calibration value
when power factor = 1.
• Calibration of power when the power factor = 0
5.
Shift the phase of the current signal channel of the phase shifter so that the
power indicated on the standard power meter shows zero.
The power value displayed on the WT210/WT230 at this point is the calibration value
when power factor = 0.
• Calibration of power with high accuracy
6.
To calibrate the WT210/WT230 with even higher accuracy, adjust the AC
voltage/current standard and phase shifter so that the display on the standard
power meter matches the calibration value of the standard power meter.
The displayed value on the WT210/WT230 at this point is the calibration value with
higher accuracy.
• Calibration at a frequency other than 60 Hz
7.
To perform calibration at a frequency other than 60 Hz, set the frequencies of
the phase shifter and the standard and carry out the procedure above in a
similar fashion. Make sure to use a standard that can guarantee adequate
accuracy against the output frequency. As a guideline, use a measuring
instrument that has three to four times the accuracy of the specifications of the
WT210/WT230.
15.2 Calibration
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...