8-4
IM 760401-01E
Recalling Measured/Computed Data (Retrieving Data from the Internal Memory)
After displaying data stored in the internal memory, you can use all display functions
and carry out integration and display these data. Furthermore, by using the
communication function, data can be output.
• Items which can be recalled
all data which can be stored.
• Aborting Recalling
The recall operation stops in the following conditions.
• when all stored data are retrieved;
• when during the recall process “oFF” is selected at the store ON/OFF setting.
Setting the Recalling Interval
Sets the time interval for repeating the recall operation.
• Setting range: 00.00.00 (0 hrs, 00 min, 00 sec) to 99.59.59 (99 hrs, 59 min, 59 sec)
• Initial value: 00.00.00
When set to 00.00.00, the recalling interval is set to the display update rate as when
the data was stored.
Recalling ON/OFF
After having set the recalling interval, select the recall menu once again. The initial
value is oFF.
• on: Recalling will start by pressing the ENTER key after selecting “on”; the
RECALL indicator will light while recalling is in progress.
• oFF: Recalling will stop by pressing the ENTER key after selecting “oFF”; the
RECALL indicator will extinguish
Note
•
During recalling, the measurement conditions/range
*
will become as those of the data being
recalled. After recalling finishes, the original measurement conditions will return.
*
Measurement range, measurement mode, measurement synchronization source, input
filter ON/OFF, scaling ON/OFF, scaling constants, averaging ON/OFF, averaging mode,
averaging coefficients, MAX hold ON/OFF, display update rate, crest factor, integration
mode, integration time, harmonic measurement function ON/OFF, PLL source, target
element, computation method of harmonic distortion factor
•
When recalling data to a personal computer by communication interface, data might be cut
due to the data length or used personal computer. In such a case, increase the recalling
interval.
8.1 Storing/Recalling Measured/Computed Data
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...