
14-36
IM 760401-01E
RECall?
Function
Queries all the settings relating to recalling data.
Syntax
RECall?
Example
RECALL? -> :RECALL:STATE 0;
INTERVAL 0,0,0
RECall:INTerval
Function
Sets the recalling interval/queries the current
setting.
Syntax
RECall:INTerval {<NRf>,<NRf>,<NRf>|
<String>}
RECall:INTerval?
{<NRf>,<NRf>,<NRf>}=0,0,0 to 99,59,59
{<String>}=HH:MM:SS
HH hour MM minutes SS seconds
Example
RECALL:INTERVAL 0,0,0
RECALL:INTERVAL “00:00:00“
RECALL:INTERVAL? -> :RECALL:
INTERVAL 0,0,0
Description If the recalling interval is set to 0 h 0 min 0 s, the
recalling interval is set to the display update rate
as when the data was stored.
RECall:PANel
Function
Retrieves the setup parameters file.
Syntax
RECall:PANel {<NRf>}
{<NRf>}=1 to 4 : file number
Example
RECALL:PANEL 1
RECall[:STATe]
Function
Turns recalling ON/OFF, queries the current
setting.
Syntax
RECall[:STATe] {<Boolean>}
RECall:STATe?
Example
RECALL:STATE ON
RECALL:STATE? -> :RECALL:STATE 1
14.3.10 RECall Group
The commands in the RECall group are used to make settings relating to, and inquires about recalling data. This
allows you to make the same settings and inquiries as can be set using the lower menus of [MEMORY]-“rECAL” or
[MEMORY]-“PnLrC”.
:RECall
:
STAT e
<Space>
OFF
ON
<NRf>
?
INTer val
<Space>
<NRf>
,
<NRf>
,
<NRf>
?
PANel
<Space>
<NRf>
?
<Character string>
;
14.3 Commands
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...