15-14
IM 760401-01E
15.4 Error Codes and Corrective Actions
Error Codes for Operation and Measurement
Error Code
Description
Corrective Action
Section
11
Received a command not used by this instrument.
Check for error in the
13.1, 14.3
command sent.
12
Parameter value specified is outside the allowed range.
Correct the value.
13.1, 14.3
13
Attempted to execute a key operation or received a
Check whether integration is in
6.4, 13.1, 14.3
communications command, while integration was running
progress or is interrupted.
or was interrupted, that cannot be executed or received
in such a state.
14
Attempted to set auto range while external sensor range
It is not possible to set auto
4.4
is selected.
range while external sensor
range is selected.
15
Attempted to execute a command or key operation that
Check whether the command
–
was protected.
or key operation is correct.
16
Attempted to execute a key operation or received a
Check whether
7.3
communications command, while harmonic measurement
harmonic measurement is in
was being performed or was interrupted, that cannot be
progress or is interrupted.
executed or received in such a state.
17
Print output time-out.
9.9
18
No data to be printed or not in the printing mode.
9.9
19
Attempted to execute a key operation or received a
–
communications command, while storing/recalling of
data being performed.
30
Invalid file data.
–
31
File is damaged.
–
32
No data stored in the internal memory.
Store data in the internal
Chapter 8
memory or select the proper file
to be stored.
33
No space to store data in the internal memory.
8.1
41
• Attempted to start integration while there is an overflow
Reset integration.
6.3
condition.
• Attempted to start integration after elapsed integration time has
reached its preset value.
42
Attempted to start integration while integration is in
6.4
progress.
43
Measurement stopped due to overflow during integration or
6.4
due to a power failure.
44
Attempted to stop integration even though integration was
6.4
not in progress.
45
Attempted to reset integration even though integration was
6.4
not in progress or integration mode was not selected.
46
Attempted to start integration while measurement of peak
6.4
overflow was inprogress or during an overrange condition.
47
Attempted to start integration in continuous integration
Set a correct preset time.
6.2
mode when integration preset time was set to “0.”
50
A/D conversion time-out.
–
51
Measurement data overflow occurred. “oL” is displayed.
2.3
52
Voltage peak overflow occurred. V OVER indicator lights up.
2.3
53
Current peak overflow occurred. A OVER indicator lights up.
2.3
54
Power factor exceeded “2” during measurement of power
5.2
factor.
55
“PFErr” was displayed at the end of power factor
5.2, 5.3
computation during measurement of phase angle.
56
Input level was too low or below measurement range during
5.4
measurement of frequency. “Err-Lo” is displayed.
57
Measured frequency exceeded the measurement range.
5.4
“Err-Hi” is displayed.
58
Computation overflow occurred. “oF” is displayed.
2.3
59
Harmonic measurement becomes “FrqEr.”
7.1
390
Overrun error.
Lower the baud rate.
11.3
Содержание WT210
Страница 1: ...Digital Power Meter IM 760401 01E 3rd Edition ...
Страница 208: ...13 18 IM 760401 01E 13 3 Sample Program Image ...
Страница 214: ...13 24 IM 760401 01E 13 5 Sample Program Output of Normal Measurement Data ...
Страница 266: ...14 50 IM 760401 01E 14 6 Sample Program Image ...
Страница 272: ...14 56 IM 760401 01E 14 8 Sample Program Output of Normal Measurement Data ...