<6. Troubleshooting>
6-10
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Operator Action
Scrolling Marquee
Main Display
Notes
Press
Press
Press
When the reset is complete, the system will
return to the measure mode. The operator will
need to re-enter the setup mode to continue
programming the setup parameters. Press
to returnto the setup mode.
PRESS TEST
VIEWENTER TO RESET
rSEt
?
This command resets
all previously set
parameters to factory
default values. Use this
command only to set
the analyzer to original
factory setup values.
6.4.1
Hard Reset
A hard reset should be performed if the analyzer becomes unresponsive or the status indicators
are flashing. Perform the following sequence:
• Carefully open the front cover of the electronics case (loosen the four corner screws).
• Press the small button located on the raised portion of the PCB to reset the electronics.
• Close the front cover of the electronics case and tighten the screws.
System settings and calibrations are preserved and analyzer returns to the measure mode. The
actual value for concentration and temperature will be displayed if the electrodes are properly
installed.
6.4.2
Serial Number and Software Revision
In the event you require technical assistance, please have the serial number available when
calling for assistance along with the software version.
• The analyzer serial number is located on the outside of enclosure.
• The software revision is accessed through the diagnostics mode.