<5. Analyzer Maintenance>
5-3
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
1. Support the bottom of the reagent bottle and release the holding clamp with one hand. Turn
the thumbscrew counterclockwise to release the reagent bottle. Unscrew the white bottle
cap and gray tubing connector assembly. Since the reagent contains concentrated acid,
which is very corrosive to skin and eyes, take the reagent to a well ventilated area such as a
laboratory fume hood. Wear rubber gloves and eye goggles. Unscrew the cap and pour the
old reagent into a suitable waste container.
Figure 5.1 Reagent Bottle Assembly
2. Take a new bottle of the reagent to a well ventilated area, such as a laboratory fume hood.
Unscrew and remove the white cap on the bottle.
3. Connect the diffusion tubing (Part number: K9705CJ) to the tubing connector assembly.
Place the tubing assembly and cap into the bottle.
4. Replace the gray cap and screw the white cap on tightly.
5. Insert the five nipples on the top of the reagent bottle into the fluid connector block.
6. Support the bottom of the reagent bottle with one hand and tighten the thumbscrew
clockwise.
7. Clamp the bottle to the fluidics panel to secure it.