<2. Analyzer Preparation>
2-5
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Power Cable
Hole Locations
Cable Glands
Electrode Cables
4 Captive Screws
Hinge Pin
Figure 2.4 Electronics Enclosure with Cable Glands
1. Open the faceplate – loosen the four screws using a Phillips head screwdriver. The
electronics faceplate will open via the hinge pin connection.
2. Remove one or two of the two unused cable glands as required for wiring power cable or
auxiliary connections. Power cable optional hole locations are shown in Figure 2.4.
3. Select and install the appropriate size cable gland or conduit fitting as required.
4. Feed the power cable through the conduit or cable glands as required.
Terminal connector
Figure 2.5 Terminal Connector Location
5. Wire the power cable to the green screw terminal connector from the options kit. Select
correct terminal for hot conductor depending on line voltage, insert ground wire into the
correct terminal and connect the other cable to the neutral terminal. Refer to Figure 2.5 for
terminal connector location.
6. Plug the terminal connector into the power supply. Refer to Figure 2.3.
7. The universal power supply uses both fuses in the fuse kit. Install by inserting the fuse in the
fuse holder and secure it using the twist and lock method. The fuses are clearly labeled with
the appropriate voltages for your system. Refer to Figure 2.3 and Figure 2.5 for the correct
fuse holder positions. Refer to the table below for fuses required.
AC Voltage
Fuse Rating
115V
200mA, 250V, Fast Acting
230V
100mA, 250V Fast Acting