<3. Analyzer Operation>
3-24
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Default
Operator Action
Scrolling Marquee
Main Display
Notes
DYN
DYN
(flashing)
SEL
SCrn
SEL CH1 or SEL
CH2 on main display
of two channel
analyzer
Press
to set custom DYN
settings for the electrode calibration
40mL/
minute
SET SAMP FLOW
RATE IN mL/M
FLO
###.#
(first digit flashing)
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits
SET SAMP FLOW
RATE IN mL/M
FLO
###.#
(change flashing
digit)
Press
to accept the setting and
advance to the next screen
2.0 ppm
SET STD
CONCENTRATION
COnC
###.#
(first digit flashing)
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits
SET STD
CONCENTRATION
COnC
###.#
(change flashing
digit)
Press
to accept the setting and
advance to the next screen
99.9 %
SET PUMP
PERCENT FLOW
FOR STD1
FLO
###.#
(first digit flashing)
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits
SET PUMP
PERCENT FLOW
FOR STD1
FLO
###.#
(change flashing
digit)
Press
to accept the setting and
advance to the next screen
100
SET PUMP RANGE
FOR STD1
rAn
100
(flashing)
Press
/
to scroll through 1,
10, 100 and 1000
SET PUMP RANGE
FOR STD1
rAn
1, 10, 100 or 1000
(flashing)
Press
to accept the setting and
advance to the next screen
52.5 %
SET PUMP
PERCENT FLOW
FOR STD2
FLO
###.#
(first digit flashing)
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits
SET PUMP
PERCENT FLOW
FOR STD2
FLO
###.#
(change flashing
digit)
Press
to accept the setting and
advance to the next screen
10
SET PUMP RANGE
FOR STD2
rAn
10
(flashing)
Press
/
to scroll through 1,
10, 100 and 1000
SET PUMP RANGE
FOR STD2
rAn
1, 10, 100 or 1000
(flashing)