<3. Analyzer Operation>
3-23
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Default
Operator Action
Scrolling Marquee
Main Display
Notes
CAL
CAL
(flashing)
SEL
SCrn
SEL CH1 or SEL
CH2 on main display
of two channel
analyzer
Press
to set the required
calibration frequency
0720
ENTER
CALIBRATION
FREQUENCY IN
HOURS
CAL
####
(first digit flashing)
To disable the
calibration alarm, set
the value to 0000
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits
ENTER
CALIBRATION
FREQUENCY IN
HOURS
CAL
####
(change flashing
digit)
Press
to accept the setting and
return to the main setup mode
CH1 or CH2
(flashing)
SEL
SCrn
Display for two
channel analyzer
only
Press
to return to the channel
specific menu options in the setup
mode
Action required for
two channel analyzer
CAL
(flashing)
SEL
SCrn
SEL CH1 or SEL
CH2 on main display
of two channel
analyzer
Press
to scroll to the next menu