6. Locate the protective cap from reference electrode kit and place it on the base of the
reference electrode to prevent it from drying out.
n
Start-Up
1. Replace the diffusion tubing if the sample flow was off for more than a few days. Tubing
becomes brittle with long-term exposure to the reagent. If the age of the reagent is not
known, replace it and note the date in the maintenance records. Re-install the reagent bottle
on the analyzer.
2. Restore sample flow to the analyzer. If necessary, adjust the pressure and flow rate through
the analyzer to 40 mL/min.
3. Remove the protective cap from the reference electrode. Re-install the electrodes on the
analyzer. Be careful not to disconnect the reservoir tubing from the reference electrode
sidearm.
4. Allow the analyzer to stabilize for approximately 1 hour and then recalibrate. Refer to
Section 4.4, Performing a DYN Calibration.