Pressure Regulator
Adjusts flow of the incoming sample stream.
Flow Valve
Used to turn off flow to the flow cell.
Flow Meter
Measures the sample flow rate.
Restrictor Tube Assembly
Used in conjunction with the pressure regulator to maintain a steady sample flow rate.
Fluid Connector Block
Connects the reagent bottle to the sample stream and flow cell assembly.
Diffusion Tubing Assembly
Semi-permeable tubing through which the reagent and iodine diffuse into the sample.
Reagent Bottle
Contains the iodine solution and acid reagent that adjusts the sample pH.
Thumbscrew
Supports the reagent bottle.
Reagent Bottle Clamp
Holds the reagent bottle securely in place.
Reagent Mixing Loop
Mixes the iodine reagent with the oxygen scavenger in the sample to form iodide, which is
measured by the iodide sensing electrode.
Flow Cell
Contains the iodide sensing electrode, reference electrode and ATC probe.
Iodide Electrode
Senses iodide ions in the sample stream and produces an electrical potential dependent on
the oxygen scavenger concentration.
Reference Electrode
Provides a constant reference potential and completes the measurement circuit.
Pressurized Reference Electrode Filling Solution Bottle
Provides a constant flow of electrolyte solution through the reference electrode for
maximum stability.
Check Valve
Prevents the backflow of electrolyte from the pressurized reservoir.
ATC Probe
Measures the sample temperature and inputs the data to the microprocessor for automatic
temperature compensation (ATC).
Standard Injection Port
Allows the connection of the dynamic calibrator tubing to the fluid connector block during a
dynamic two point calibration.
Air Pump
Provides air to pressurize the filling solution bottle.
LCD Display
Provides digital readouts of concentration, temperature, millivolts and error codes.
Keypad
Consists of five mode keys, four prompt indicator lights, two scroll keys and one key for
entering data. Mode and error indicators are also incorporated on keypad.
Status Indicator
Two LED lights that illuminate according to current status of the analyzer.
Green Light: Indicates that system is in correct working condition.
Yellow Light: Indicates a warning, system in hold or that maintenance is required.
Red Light:
Indicates that something is seriously wrong.
NOTE
When either the yellow or red LED is lit, there may be an entry in the diagnostics mode that
indicates the error. The logging feature must be initiated in the setup mode. Refer to Section 3.1,
Use of the Setup Mode for instructions.