<4. Calibration>
4-3
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
4.3
Dynamic Calibrator and Syringe Setup
1. Connect the calibration tubing to the syringe using the luer adapter and the 1/4’’ female
connector. Only hand tighten the components, do not overtighten them.
2. Loosen the blue knob on the dynamic calibrator by turning it approximately 1/4 of a turn
clockwise and pressing up on the bottom, inside part of the blue knob. Lift the blue syringe
holder and turn it away from the syringe cradle. Set the filled syringe in the center syringe
cradle. Rotate the syringe holder back in place across the syringe. Gently tighten the blue
knob by turning it 1/4 of a turn counterclockwise.
3. Lift the black drive carriage straight up. Move the carriage back and re-engage it with the
gears, leaving room for the syringe. Check the alignment of the drive carriage to ensure it is
perfectly straight.
4. Replace the drive carriage on the gears. Advance the carriage to meet the syringe by
momentarily turning the pump on high speed.
NOTE
Do no attempt to push the carriage manually. This may result in damage to the pump
mechanism.
5. Set the blue endpoint limiter midway between the white indicator marks.
4.4
Performing a DYN Calibration
• To begin a DYN calibration, the system starts from the measure mode.
• The analyzer must be set to the hydrazine application in the setup mode prior to performing
a calibration. Refer to Section 3.1, Use of the Setup Mode.
NOTE
A three-way valve can be installed upstream of the analyzer inlet to allow easy switching to a
oxygen scavenger-free sample such as demineralizer output or condensate pump discharge.
Step
Operator Action
Scrolling Marquee
Main Display
Notes
1
Make sure that the analyzer is in the
measure mode.
Actual temperature
reading
Actual
concentration
reading
2
Press
to initiate the calibration.
CAL appears as
mode
3
If the calibration password is active,
enter the calibration password by
pressing
/
to set the first
digit,
to move to the next digit,
/
to set the values of the
remaining digits and
to move
through the remaining digits.
Press
to accept the password.
ENTER PASSWORD
PASS
# # # #
4
CH1 or CH2
(flashing)
CAL
CHnL
Displayed for two
channel analyzer
5
Press
/
to select the channel
that will be calibrated and press
to
begin the calibration.
CH1 or CH2
(flashing)
CAL
CHnL
Action required for
two channel analyzer
6
DYN or OFFL
(flashing)
CAL
tYPE
7
Press
/
to select DYN and
press
to accept the setting.
DYN
(flashing)
CAL
tYPE