<2. Analyzer Preparation>
2-10
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
CAUTION
Failure to pressurize the filling solution bottle will lead to noisy and drifting output signals.
9. Plug the electrode cable marked “Reference” into the top of the electrode, and tighten the
screw cap. Be sure to push back the black cap to verify a secure connection between the
male and female pin connection prior to tightening.
10. Tighten the screw cap connection to the cable.
NOTE
Do not twist the cable while tightening the connection. Twisting may cause damage requiring
premature replacement of the cable.
11. Insert the reference electrode into its port in the flow cell cap. Refer Figure 2.8 for the
location.
Bottle
Clip (on Panel)
Figure 2.10 Reference Mounting Clip