<3. Analyzer Operation>
3-21
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Default
Operator Action
Scrolling Marquee
Main Display
Notes
TEST
TEST
(flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of two
channel analyzer
Press
to test the mA outputs (4-
20) and relays (rLY)
To Test OUT2
4-20
Press
/
to scroll between
4-20 and rLY settings and select 4-20
SELECT mA OR
RELAY
tESt 4 - 20
(flashing)
Press
to test 4-20 outputs
SELECT mA1 OR
mA2 OUTPUT
4 - 20
OUT1 or OUT2
(flashing)
Press
/
to scroll between
OUT1 and OUT2 settings and select
OUT2
SELECT mA1 OR
mA2 OUTPUT
4 - 20
OUT1 or OUT2
(flashing)
Press
to display the OUT2 low
Actual
low value
displayed
mA TEMP OUTPUT
LOW VALUE
LO
#.# c
4.0 mA are sourced at
output terminal 11
Press
to display the OUT2 high
Actual
high value
displayed
mA TEMP OUTPUT
HIGH VALUE
HI
#.# c
20 mA are sourced at
output terminal 11
Press
to accept the test and
return to the main setup mod
CH1 or CH2
(flashing)
SEL
SCrn
Display for two
channel analyzer only
Press
to return to the channel
specific menu options in the setup
mode
Action required for two
channel analyzer only
TEST
(flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of two
channel analyzer
Press
to test the mA outputs (4-
20) and relays (rLY)
To Test mA1:
4-20
Press
/
to scroll between
4-20 and rLY settings and select 4-20
SELECT mA OR
RELAY
tESt 4 - 20
(flashing)
Press
to test 4-20 outputs
SELECT mA1 OR
mA2 OUTPUT
4 - 20
OUT1 or OUT2
(flashing)
Press
/
to scroll between
OUT1 and OUT2 settings and select
OUT1
SELECT mA1 OR
mA2 OUTPUT
4 - 20
OUT1
(flashing)
Press
to display the mA1 low
value
Actual
low value
displayed
mA SENSOR
OUTPUT LOW
VALUE
LO
#.#
4.0 mA are sourced at
output terminal 9
Press
to display the mA1 high
Actual
high value
displayed
mA SENSOR
OUTPUT HIGH
VALUE
HI
#.#
20.0 mA are sourced
at output terminal 9
Press
to accept the test and
return to the main setup mode
CH1 or CH2
(flashing)
SEL
SCrn
Display for two
channel analyzer only
Press
to return to the channel
specific menu options in the setup
mode
Action required for two
channel analyzer only