The offline calibration is essentially a one point calibration. To perform an offline calibration, a
sample is taken from the bypass of the analyzer; the sample concentration value is stored in
memory; the sample is analyzed by an alternate method of choice; the previously stored reading
is adjusted to the lab method result; and the analyzer is then returned to the analysis mode. The
term “offline calibration” refers only to the fact that a sample from WE430 analyzer bypass is
taken “offline” for laboratory analysis; in fact, no downtime is experienced during the procedure
and the analyzer remains online throughout.
1.3
Fluidics Diagram
Figure 1.2 Fluidics Diagram
1.4
Glossary
Refer to Figure 1.2.
Inlet Valve
Accepts the sample stream via 1/4 inch NPTF connector. The operator must supply the
sample with a pressure between 8 and 100 psig.
Inlet Filter
60 micron stainless steel filter traps particulate matter in the sample stream.
Bypass/Needle Valve
Used to redirect flow in the bypass system.