<4. Calibration>
4-6
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Step
Operator Action
Scrolling Marquee Main Display
Notes
Error
110
Press
to continue the calibration
despite the bad slope.
or
Press
to abort the calibration and
refer to Exit 3.
CH2 BAD SLOPE
PRESS ENTER
E110
The calculated slope
was outside of the
recommended range.
Perform electrode
cleaning and
conditioning, refer to
the troubleshooting
section and repeat the
calibration.
Error
125
Press
to abort the calibration and
refer to Exit 3.
Verify that the correct standards were
used during the calibration, the standard
values were entered correctly and in the
right order and the electrodes are working
properly and then repeat the calibration.
CH1 BAD
OFFSETPRESS
ENTER
E125
Use new standards.
Perform electrode
cleaning, conditioning
and troubleshooting.
Error
126
Press
to abort the calibration and
refer to Exit 3.
Verify that the correct standards were
used during the calibration, the standard
values were entered correctly and in the
right order and the electrodes are working
properly and then repeat the calibration.
CH2 BAD
OFFSETPRESS
ENTER
E126
Use new buffers or
standards. Perform
electrode cleaning,
conditioning and
troubleshooting.
4.6
Calibration At Custom Concentrations
Using DYN
The WE430 Hydrazine analyzer software has default values to accommodate standards added
using the dynamic calibrator. The operator has the option to use alternate standards by changing,
in the setup mode, the default values.
Hydrazine Default Values
Analyzer Flow Rate
= 40 mL/minute
Standard Concentration
= 2 ppm
Standard 1 Percent Flow
= 99.9%
Standard 1 Range
= x/100
Standard 2 Percent Flow
= 52.5%
Standard 2 Range
= x/10
Dynamic Calibrator Settings
Settings on the dynamic calibrator are based on a 20 cc syringe at 100% (39.7 mL/minute). The
flow rate of the standard (F
std
) is the 39.7 mL/minute multiplied by the percent setting, divided by
the range. Refer to the dynamic calibrator user guide for additional information.
C = [(C
b
• F
b
) + (C
std
• F
std
)] / (F
b
+ F
std
)
Where:
C = Measured concentration during calibration
C
b
= Background concentration (assumed to be zero)
C
std
= Concentration of standard in the syringe
F
b
= Background flow rate of analyzer (approximately 40 mL/min).
F
std
= Flow rate of standard