<5. Parameter setting>
5-18
IM 12J05D02-01E
Code Display
Function
Function detail
X Y Z Default values
mA Output Functions
30
*mA
mA output
Output 1 is 0- 20 mA
0
1.1
Output 1 is 4- 20 mA
1
Output 2 is 0- 20 mA
0
Output 2 is 4- 20 mA
1
31
*OUTP.F
*D/R
Output function
Control action
Process value on mA1
0
0.2
Table on mA1
1
Process value on mA2
0
Table on mA2
1
Temperature on mA2
2
PI control on mA2
3
Reverse acting PI-control
0
0
Direct acting PI control
1
32
*BURN
Burn-out
No burn-out function on mA1
0
0.0
Burn low on fail on mA1
1
Burn high on fail on mA1
2
No burn-out function on mA2
0
Burn low on fail on mA2
1
Burn high on fail on mA2
2
33
*RGmA2 Prop. range
Proportional band: 0- 50 mg/L
5.0 mg/L
0-300 % saturation
34
*tImA2
Integral time
Integral time in seconds 10 - 1999 s
100 s
35
*TABL1
Table for mA1
21 point table: Process value at
0, 5, .. 100 % of span
none
36
*TABL2
Table for mA2
21 point table: 0, 5, .. 100 %
none
37
*DAMP
Output damping
damping 0- 120 s
0 s
38
*SWTCH Range 1B
Disabled
0
0
Enabled
1
39
Not used
Note To enable wash function, set Service Code 38 to 0, Service Code 42 to 4.0.0, and Service
Code 51 to 0.1.0.
5.3.5
Contact outputs
Code 40, *S1 & *S2 Process relays can be set for a variety of alarm and control function.
41, 42 & 43
Digit “X” sets the type of trigger:
Off means that the relay is not active
Low setpoint means that the relay is triggered by a decreasing
measurement.
High setpoint means that the relay is triggered by an increasing
measurement “HOLD” active means that there is maintenance activity in
progress so the measurement is not live.
For *S3
There is the extra possibility to set up for a wash controller.
For *S4
There is the extra possibility to set up for “FAIL” indication.
Digit “Y” sets the control action:
Process alarm is a simple On/Off trip controlled by the high/low setpoint.
Proportional duty cycle control has a pulse width modulation for
proportional dosing with solenoid valves.
Proportional frequency control is used for controlling electrically
positioned valves.
Temperature alarm is an On/Off trip on the measured temperature.
Digit “Z” sets the control parameter:
With or without “integral” action.
9th Edition: Aug. 05, 2015-00