<5. Parameter setting>
5-23
IM 12J05D02-01E
Code Display
Function
Function detail
X Y Z Default values
User Interface
50
*RET
Auto return
No return to meas. from HOLD
0
1
Return to meas. after 10 min
1
51
*MODE
add. to MAINT
Setpoint adj. disabled
0
0.0.0
Setpoint adj. also in MAINT
1
Wash start disabled
0
Wash start in maintenance mode
1
Manual pressure disabled
0
Manual pressure enabled
1
52
*PASS
Password
Pass: 111 for #=1, 333
for #=2, 777 for #=3,
888 for #=4, 123 for
#=5, 957 for #=6, 331
for #=7, 546 for #=8,
847 for #=9
Direct access to MAINT
0
0.0.0
MAINT protected by password
#
Direct access to COMM
0
COMM protected by password
#
Direct access to SERV
0
SERV protected by password
#
53
*Err.01
*Err.02
*Err.03
*Err.04
*Err.07
*Err.08
*Err.09
*Err.12
*Err.16
*Err.22
Error setting
Response & Time failure
Soft fail
Response & Time failure
Hard fail
Zero calibration
Soft fail
Zero calibration
Hard fail
Slope calibration
Soft fail
Slope calibration
Hard fail
Membrane failure
Soft fail
Membrane failure
Hard fail
Temperature sensor open
Soft fail
Temperature sensor open
Hard fail
Temp. sensor shorted
Soft fail
Temp. sensor shorted
Hard fail
Cell current abnormal
Soft fail
Cell current abnormal
Hard fail
Abnormal measured value Soft fail
Abnormal measured value Hard fail
Cell for maintenance
Soft fail
interval time exceeded
Cell for maintenance
Hard fail
interval time exceeded
Control time-out
Soft fail
Control time-out
Hard fail
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
1 Hard
1 Hard
1 Hard
1 Hard
1 Hard
1 Hard
1 Hard
1 Hard
0 Soft
0 Soft
*SOFT
LCD Fail Constant
0
0 LCD Fail
LCD only
1
54
*E12
overrange limit
E12 active at 0 - 300 % sat (Note)
150%
55
*CALL.M E16 setting
No call for maintenance
0
0
Call for maint. enabled
1
E16 message after 1- 250 days
30
56
*UNIT
PV unit
Unit of measurement mg/L
0
0
Unit of measurement ppm
1
Unit of measurement % sat.
2
57-
59
Not used
Note: For setting a value greater than 199.9, move the decimal point. For a value greater than 200%, decimal places cannot be
used.
5.3.7
Communication setup
Code 60 *COMM.
The settings should be adjusted to suit the communicating device
connected to the RS-485 port.
*SET.
*ADDR.
For the Yokogawa PC402 software package, the default settings match
the software as shipped.
9th Edition: Aug. 05, 2015-00
Содержание EXA DO402G
Страница 9: ...Blank Page...
Страница 23: ...Blank Page...
Страница 77: ...Blank Page...
Страница 81: ...Blank Page...
Страница 85: ...Blank Page...
Страница 86: ...9 Spare Parts 9 1 IM 12J05D02 01E 9 Spare Parts See Customer Maintenance Parts List 9th Edition Aug 05 2015 00...
Страница 87: ...Blank Page...
Страница 95: ...Blank Page...