<3. Installation and Wiring>
3-13
IM 12J05D02-01E
The overall shield of the cable is marked with 14 and must be connected to terminal 14.
CAUTION
A jumper cable is placed to connect converter terminals 13 and 17.
Make sure that connecting the sensor cable to the IE pin or pin 13 or pulling the sensor cable
does not cause the jumper cable to come loose. Insufficient tightening may cause unstable or
false measurements.
F0311.ai
Digital
Commnucations
12
13
15
16
63 66 65 62 61 95 94 93 92 91
SCREEN
mA2
mA1
SCREEN
TL
TL
17 11
CONT
mA OUTPUT
14
22
21
18
SCREEN
99
SCREEN 2
23
mA Output
(IE)
(T1) (T2)
(RE)
(LE)
Short
Sensor Input
Contact
Input
Shield
Temp.
Liquid Earth
Wirning galvanic Sensor
Figure 3.12 Wiring galvanic sensor
3.7
Wiring other galvanic sensors
Consult the users manual for the color identification of the sensor cable and connect temperature
compensator, cathode and anode to the terminals: 11, 12, 13 and 15 as described above.
Connect the cable shield to 14 if there is one available. Sensor diagnostics is normally not
possible.
Connect any of the following temperature compensators: PT100, PT1000, PB36, 22kNTC
Connect only to the cathode and anode of galvanic sensors.
Do not connect to other devices.
CAUTION
A jumper cable is placed to connect converter terminals 13 and 17.
Make sure that connecting the sensor cable to the IE pin or pin 13 or pulling the sensor cable
does not cause the jumper cable to come loose. Insufficient tightening may cause unstable or
false measurements.
9th Edition: Aug. 05, 2015-00