<7. Maintenance>
7-1
IM 12J05D02-01E
7. Maintenance
It is important for maintaining the measurement accuracy of the EXA DO series of wire
dissolvedoxygen metering system to perform inspection and maintenance at fixed intervals. It
also serves to prevent problems from arising.
This chapter describes daily inspection and maintenance for the purpose of maintaining system
performance.
7.1
Overall dissolved oxygen metering system
Tables 7.1 and 7.2 show the inspection and maintenance items for equipment composing 4-wire
dissolved-oxygen metering system. The procedure of inspection and maintenance for the
dissolved oxygen converter is described in Section 7.2. For inspection and maintenance of the
sensors and holders, see each user’s manual respectively.
7.1.1
Inspection and maintenance to be implemented
periodically
Table 7.1 shows the items which are recommended to be inspected and maintained periodically.
Table 7.1 Items for periodic inspection and maintenance of 4-wire dissolved-oxygen metering system
Instrument or apparatus
Item
Recommended Interval
All equipment
a. Calibration
a. 1 to 2 months
Dissolved oxygen sensor
a. Clean the tip of the sensor
b. Polishing of (silver) electrode surface
and replacement of electrolyte
c. Replacement of membrane assembly
d. Replacement of sensor cap
a. 1 to 2 months
b. 6 to 8 months
c. 6 to 8 months
d. 6 to 12 months
Dissolved oxygen converter a. Inspection of temperature indication
(error check)
b. Replacement of fuse
a. 6 months
b. 1 to 2 years
Float holder
a. Sweeping of winch mechanism
a. (depends on installation
environment)
Immersion holder
a. Inspection of sensor holding O-ring
b. Sweeping of cleaner nozzle
c. Inspection for leakage from the
solenoid valve for cleaning utility
a. 6 months
b. 1 to 2 months
c. 6 months
7.1.2
Inspection and maintenance to be implemented on
occasion
Occasionally implement the maintenance that is not directly related to measurement functions
and maintenance associated with errors.
Table 7.2 Items for occasional inspection and maintenance of 4-wire dissolved-oxygen metering
Faulty phenomenon
System content of inspection
Error E2 occurrence
Check whether the dissolved-oxygen sensor membrane is contaminated or
not.
Check whether the nozzle is clogged or not.
Error E3 occurrence
Polish the silver electrode surface of the galvanic dissolvedoxygen sensor,
platinum surface of the polarographic dissolvedoxygen sensor, and replace
the electrolyte. Check that the connection between sensor and sensor
cable is fit, and LED emission of the optical dissolved-oxygen sensor.
It is difficult to see the inside of
the converter through the window
Wipe off the dirt on the transparent window using tissue paper or the like.
(Use detergent for hard-to-clean dirt.)
9th Edition: Aug. 05, 2015-00
Содержание EXA DO402G
Страница 9: ...Blank Page...
Страница 23: ...Blank Page...
Страница 77: ...Blank Page...
Страница 81: ...Blank Page...
Страница 85: ...Blank Page...
Страница 86: ...9 Spare Parts 9 1 IM 12J05D02 01E 9 Spare Parts See Customer Maintenance Parts List 9th Edition Aug 05 2015 00...
Страница 87: ...Blank Page...
Страница 95: ...Blank Page...