<3. Installation and Wiring>
3-12
IM 12J05D02-01E
3.5.2
Analog output signals
The output signals consist of active current signals of either 0-20 mA or 4-20 mA. The maximum
load can be 600 ohms on each.
It should be necessary to use screening/shielding on the output signal cables. Terminal 63 is
used to connect the shielding.
Figure 3.11
Cable gland assembly
3.6
Wiring the standard galvanic sensor
Do not use cables other than the dedicated cables.
Tighten the cables to 0.5 N•m torque.
The sensor cable has markers on the individual wires.
These markings refer to the markers on the terminals.
The temperature compensator has two wires with the markings T1, T2 and/or 11, 12 and must be
connected to terminal 11 and 12.
The measuring electrode: the cathode is marked with IE and/or 13 and must be connected to
terminal 13.
The reference electrode: the anode is marked with RE and/or 15 and must be connected to
terminal 15.
The liquid earth ( solution ground) is marked with 16 and must be connected to terminal 16.
CAUTION
The liquid earth (solution ground) connection is valid only when the floating ball holder, Model
PB350G or PB360G, is used. Otherwise, do not make a connection to terminal 16. The
connection, if made, may result in abnormal readings. Even if the floating ball holder is used,
when the liquid earth (solution ground) connection is not used, then disconnect the wiring to
terminal 16. (The disconnected wire should be covered with an insulating tube to avoid contact
with other terminals.)
9th Edition: Aug. 05, 2015-00
Содержание EXA DO402G
Страница 9: ...Blank Page...
Страница 23: ...Blank Page...
Страница 77: ...Blank Page...
Страница 81: ...Blank Page...
Страница 85: ...Blank Page...
Страница 86: ...9 Spare Parts 9 1 IM 12J05D02 01E 9 Spare Parts See Customer Maintenance Parts List 9th Edition Aug 05 2015 00...
Страница 87: ...Blank Page...
Страница 95: ...Blank Page...