<8. Troubleshooting>
8-1
IM 12J05D02-01E
8. Troubleshooting
This chapter describes the countermeasures for failures, classifying the cases into three
categories: dissolvedoxygen converter failure, detection of failure with the self-diagnosis function,
and abnormal measured values. The causes for abnormal measured values are not limited to
equipment failures. If an abnormal phenomenon occurs, first check the following items:
• Is the property of the measuring solution different from normal ?
• Is the dissolvedoxygen sensor properly installed?
8.1
Measures in the case of converter operation
failure
8.1.1
No dissolved-oxygen converter operation
If the converter does not operate even when the power is supplied, one cause may be a blown
fuse inside the converter.
Examine the fuse by turning off the power supply (see Subsection 3.3.1 on page 3.5).
If the fuse is blown, replace it with a new one. If fuses frequently blow and the cause is not clear,
request an inspection by Yokogawa.
If the fuse is normal, examine the wiring system.
8.1.2
Operation key or display failure
If the operation keys do not operate smoothly or the display fails (e.g., a missing character
segment), repair of the printed circuit board (digital board) (replacement with a new one) is
required.
After the printed circuit board is replaced, operation checks and parameter settings are
necessary.
Contact Yokogawa and request board-replacement work.
8.2
Measures in the case of failure (Error)
detection
If a failure is detected through the self-diagnosis of the DO402G dissolved oxygen converter,
the FAIL contact is closed. The FAIL lamp on the operation panel lights up and an error number
appears in the data display.
NOTE
If an error is detected during configuration, the FAIL contact signal is output immediately but the
error number is displayed after that action or operation is completed.
When a FAIL contact signal is output, take measures according to Table 8.1.
9th Edition: Aug. 05, 2015-00
Содержание EXA DO402G
Страница 9: ...Blank Page...
Страница 23: ...Blank Page...
Страница 77: ...Blank Page...
Страница 81: ...Blank Page...
Страница 85: ...Blank Page...
Страница 86: ...9 Spare Parts 9 1 IM 12J05D02 01E 9 Spare Parts See Customer Maintenance Parts List 9th Edition Aug 05 2015 00...
Страница 87: ...Blank Page...
Страница 95: ...Blank Page...