<10. Appendix>
10-4
IM 12J05D02-01E
Code
Error description
Possible cause
Suggested remedy
E16
Call for maintenance inetrval time
exceeded.
System not maintained in preset
time period
Perfrom maintenance
Reset interval
E17
Output span too small
Incorrect configuration by user
Reprogram
E18
Table values make no sense
Wrong data programmed
Reprogram
E19
Programmed values outside acceptable
limits
Incorrect configuration by user
Reprogram
E20
All programmed data lost
Fault in electronics
Very severe interference
Contact Yokogawa
E21
Checksum error
Software problem
Contact Yokogawa
E22
Alarm activation time exceeded
Process control not effective
within set time
Check control equipment
Adjust value in code 47
9th Edition: Aug. 05, 2015-00