<6. Calibration Procesure>
6-5
IM 12J05D02-01E
6.3.3
Procedure for Water calibration
YES NO
YES NO
YES NO
YES
The instruments waits for
the reading to stabilize.
(The display flashes)
When reading Is stable,
the CAL.END message
appears.
NO
YES
NO
YES
YES
NO
YES
MEASURE
MAN.CAL
DISPLAY
HOLD
NO
MODE
YES
ENT
YOKOGAWA
MODE
AIR.CAL
CONTACTS
S1
S2
WASH/S3
FAIL/S4
H2O.CAL
NO
YES
YES
Place sensor
in calibration
solution
Press
YES
for single point
adjustment.
Note: To start calibration
with a zero solution,
press
NO
(zero
calibration must be
activated in service
code 21).
Transfer to zero
solution and press
YES
.
When the reading is stable, the CAL END
message appears. Press
YES
.
After briefly displaying WAIT, the display returns to
the normal readout.
After briefly displaying WAIT, the display
returns to the normal readout.
NO
to *2
to *2
from *2
to *1
from *1
NO
NO
NO
NO
If “X.X.1” is selected in Service
Code 51,
the M.PR.ON message will appear
to indicate that the manual
pressure is enabled. Press
YES
to
proceed.
NO
YES
to *2
to *2
NO
9th Edition: Aug. 05, 2015-00