<7. Maintenance>
7-3
IM 12J05D02-01E
l
230 V AC version
Rated breaking current: 35 A or 10 times the rated current, whichever is greater (low breaking)
Maximum rated voltage: 250 V
Maximum rated current: 100 mA
Compliance:
UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law
Part number:
A1103EF
l
How to replace the fuse
1. Before replacing the fuse, turn off power to the instrument at the external breaker.
2. Remove the instrument cover and the high-voltage shield plate.
3. Remove the cover from the fuse holder by pulling out by hand.
4. Remove the fuse and install a new, recommended or equivalent fuse on the holder.
5. Put the fuse cover back on the holder securely.
6. Install the high-voltage shield plate.
It is recommended that the fuse be replaced every two years even if it has not blown.
Shield Plate
Fuse Cover
Fuse Holder
Figure 7.1 How to replace the fuse
9th Edition: Aug. 05, 2015-00