<4. Operation; Display Functions And Setting>
4-1
IM 12J05D02-01E
4.
Operation; Display Functions And
Setting
4.1
Operator interface
This section provides an overview of the operation of the EXA operator interface. The basic
procedures for obtaining access to the three levels of operation are described briefly. For a step-
by-step guide to data entry, refer to the relevant section of this instruction manual. Figure 4.1
shows the EXA operator interface.
l
LEVEL 1: Maintenance
These functions are accessible by pushbutton through a flexible front cover window. The
functions make up the normal day-to-day operations that an operator may be required to
complete. Adjustment of the display and routine calibration are among the features accessible in
this way. (See table 4.1).
l
LEVEL 2: Commissioning
A second menu is exposed when the EXA front cover is removed and the display board is
revealed. Users gain access to this menu by pressing the button marked * in the lower right of
the display board. This menu is used to set such values as the output ranges and hold and wash
features. It also gives access to the service menu. (See table 4.1).
l
LEVEL 3: Service
For more advanced configuration selections, press the button marked * , then press “NO”
repeatedly until you reach *SERV. Now push the “YES” button. Selecting and entering “Service
Code” numbers in the commissioning menu provide access to the more advanced functions. An
explanation of the Service Codes is listed in chapter 5 and an overview table is shown in chapter
10.
Table 4.1 Operations overview
Routine
Function
Chapter
Maintenance
(level 1)
AIR.CAL
H2O.CAL
MAN.CAL
DISP
HOLD
SETP
WASH
M.TEMP
M.PRESS
Calibration in air (preferred method)
Calibration in aerated water
Process calibration based on grab sample
Read auxiliary data and set message display
Switch HOLD on/off *
Adjust alarm setpoints *
Start wash cycle manually *
Adjust manual temp *
Adjust manual press
6
6
6
4
5
5
5
5
5
Commissioning
(level 2)
*SETP
*RANGE
*HOLD
*WASH
Adjust alarm setpoints
Adjust output range(s)
Activate HOLD function
Activate and configure the wash timer
5
5
5
5
Service
(level 3)
*SERV
Configure the specialized functions of the
converter
5
*: Only possible if activated
NOTE: All three levels may be separately protected by a password. See Service Code 52 in
chapter 5 Service Code table for details on setting passwords.
9th Edition: Aug. 05, 2015-00
Содержание EXA DO402G
Страница 9: ...Blank Page...
Страница 23: ...Blank Page...
Страница 77: ...Blank Page...
Страница 81: ...Blank Page...
Страница 85: ...Blank Page...
Страница 86: ...9 Spare Parts 9 1 IM 12J05D02 01E 9 Spare Parts See Customer Maintenance Parts List 9th Edition Aug 05 2015 00...
Страница 87: ...Blank Page...
Страница 95: ...Blank Page...