<5. Parameter setting>
5-5
IM 12J05D02-01E
5.2
Commissioning mode
In order to obtain peak performance from the EXA converter, you must set it up for each custom
application. Refer to section 1-4 for standard configurations and options.
*SETP
Alarms are set by default
S1 - high process alarm
S2 - low process alarm
S3 - WASH
S4 - FAIL
The setpoints are at arbitrary default value. Therefore, you must set these
to meaningful values, or set them to off. (See service codes 40 to 49 and
user interface codes 50 to 59).
*RANGE
mA output 1 is set as default to 0 - 20 mg/L
mA output 2 is set as default to 0 - 50 °C
For enhanced resolution in more stable measuring processes, it may be
desirable to select 0 - 5 mg/L range, for example, and maybe 0 - 25 °C
temperature range.
Service codes 30 to 39 can be used to choose other output parameters
on mA output 2.
*HOLD
The EXA converter has the ability to “hold” the output during maintenance
periods. This parameter should be set up to hold the last measured value,
or a fixed value to suit the process.
*WASH
The EXA can be set up to control a wash cleaner. When using this
function, the timings must be configured for interval, wash and recovery
periods.
*SERV
This selection provides access to the service menu.
What follows are pictorial descriptions of typical frontplate pushbutton sequences for each
parameter setting function. By following the simple YES/NO prompts and arrow keys, users can
navigate through the process of setting range, setpoints, hold, wash and service functions.
9th Edition: Aug. 05, 2015-00
Содержание EXA DO402G
Страница 9: ...Blank Page...
Страница 23: ...Blank Page...
Страница 77: ...Blank Page...
Страница 81: ...Blank Page...
Страница 85: ...Blank Page...
Страница 86: ...9 Spare Parts 9 1 IM 12J05D02 01E 9 Spare Parts See Customer Maintenance Parts List 9th Edition Aug 05 2015 00...
Страница 87: ...Blank Page...
Страница 95: ...Blank Page...