<10. Appendix>
10-2
IM 12J05D02-01E
10.4 Wash
Variable
Default
Lower limit
Upper limit
Interval
6 hours
0.1 hours
36.0 hours
Wash time
0.5 minutes
0.1 minutes
10.0 minutes
Recovery time
0.5 minutes
0.1 minutes
10.0 minutes
10.5 User setting table
FUNCTION
SETTING DEFAULTS
USER SETTINGS
Parameter specific functions
01
*S.TYPE 0
Galvanic (7 μA)
*I.CELL
0
3.75 μA
02
*CHECK 0.1.0
Zero Off
Slope On
Membrane Off
04
*SAL.TY 0
Off
Temperature measuring functions
10
*T.SENS 1
Pt1000
11
*T.UNIT
0
°C
12
*T.ADJ
None
13
*T.MAN
0
Off
Calibration parameter functions
20
*ΔT.SEC 60
sec.
*Δmg/L
0.05
mg/L
21
*0.CAL
0
Off
22
*ZERO
0
μA (nA)
*SLOPE
3.75 (50)
μA (nA)
mA outputs
30
*mA
1.1
both 4-20mA
31
*OUTP.F 0.2
DO & Temp.
*D/R
0
Reverse (control)
32
*BURN
0.0
both off
33
*RG.mA2 5.0 mg/L
only for PI control
34
*tl.mA2
100 sec
only for PI control
35
*TABL1
21 pt table see code 31, §10-1
36
*TABL2
21 pt table see code 31, §10-1
37
*DAMP
0
sec.
38
*SWTCH 0
Off
Contacts
40
*S1
2.0.0
high process Al.
41
*S2
1.0.0
low process Al.
42
*S3
4.0.0
WASH
43
*S4
4.0.0
FAIL
44
*D.TIME
0.2
sec.
*P.HYST 0.1
mg/L
*T.HYST 0.5
°C
45
*RANGE 0.5
mg/L
*PER
10.0
sec.
*FREQ
70
p/min
46
*tI.CNT
100 sec.
only for PI control
47
*EXPIR
0
off
*tE.min
15
min
User Interface
50
*RET
1
on
9th Edition: Aug. 05, 2015-00