6-16
IM AQ1200-01EN
6.8 Displaying Differential Traces
Procedure
Differential Trace Screen
1.
Use the
rotary knob
and
ENTER
to select
OTDR
.
2
.
Press the
Advanced Analysis
soft key and
then the
Diff. Trace
soft key. The following screen
appears.
When the suffix code is -HE, press the
Analysis
soft key, the
Advanced Analysis
soft key and then
Diff. Trace
soft key.
Selects the current trace
(Trace 1, Trace 2)
Executes differential trace analysis
The differential trace analysis screen appears.
Load traces to display in Trace 1 and Trace 2.
The file operation screen appears
(see section 17.2).
When marker mode is set to Marker
When marker mode is set to Line
Explanation
Loading Trace Data
The type of file that you can load is SOR.
Note
A differential trace of traces that meet the following conditions can be displayed.
• Sample interval
• Measurement start position
Current Trace
The current trace can be set to one of the traces that are used for taking the difference.
Cursors are displayed on the current trace. Return loss and splice loss that are displayed in the
measurement conditions are those of the current trace.