3-12
IM AQ1200-01EN
3.6 Setting Analysis Conditions
<Detail Mode and Multi WL Mode>
Procedure
Analysis Setup Screen
1.
Using the
rotary knob
and
ENTER
, select
OTDR
.
2.
Press
SETUP
and then the
Analysis Setup
soft key.
3.
Using the
rotary knob
and
ENTER
, set Mode to
Detail
or
Multi WL
to display the following
screen.
Detail Mode
Set the fault event to On or Off.
If you set it to On, the following screen appears.
Set Mode to “Detail.”
Displays the specified wavelength
Set the index of refraction (1.30000 to 1.79999).
Set the backscattering light level (–10.00dB to –64.99dB).
Set the approximation method (marker; LSA, TPA).
Set the approximation method (event; LSA, TPA).
Set the event’ s splice loss (0.01dB to 9.99dB).
Set the event’ s return loss (20dB to 70dB).
Set the end of fiber (Fresnel reflection; 3dB to 65dB).
Turns splitter detection on or off
Threshold for splitter detection
Set the fault event’s splice loss (0.01dB to 9.99dB).
Set the fault event’s return loss (20dB to 70dB).
Set the fault event’s PASS/FAIL judgement display
(Display, Hide).
Initializes the settings
Press to reset the settings to
their factory defaults.
Multi WL Mode
For each wavelength set in the measurement conditions, set the index of refraction and backscatter
level. In addition to the settings for Detail mode, you can set macro bending (bending loss).
Set Mode to “Multi WL.”
Set macro bending to On or Off.
If you set it to On, the following screen appears.
Turns the macro bending (bending loss)
display on or off
Set the macro bending threshold
(0.001dB to 99.999dB).