20-8
IM AQ1200-01EN
20.5 Updating the Firmware
Procedure
Version Up Screen
Press
SETUP
and then the
Version Up
soft key to display the following screen.
Using the
rotary knob
and
ENTER
, select
the firmware update file.
Executes the firmware update
Select the drive that contains the
firmware update file (internal memory,
USB memory).
Explanation
To update the firmware, select a firmware update file (.YMC extension).
• You can change file names, except for the extension, as necessary. For instructions on how to
change the file name, see section 18.4.
• When the firmware update ends successfully, the Instrument restarts automatically.
• When the firmware update fails, “Cannot load this file. Invalid file ∙∙∙∙∙∙” appears. Make sure that
there are no problems with the firmware update file.
• For instructions on how to view the version of your Instrument, see section 20.3.
• Connect the AC adapter.
• You cannot revert to firmware version earlier than 3.02.