11.1 Making Preparations and Adjustments for Loss
Testing
Procedure
Performing Zero Set
Remove the optical fiber cables from the Instrument and close the optical connector covers, or make
sure that the power meter is not receiving any light, and then start the optical power meter zero set
procedure.
1.
Using the
rotary knob
and
ENTER
, select
Auto Loss Test
.
2.
Press
SETUP
and then the
OPM Setup
soft key to display the following screen.
Performs zero set
Adjusting the Optical Power
WARNING
• While the Instrument is producing light, do not remove the optical fiber cable, because light
is emitted from the light source port. Visual impairment may occur if the light enters the
eye.
• Close the covers of any light source ports that do not have optical fiber cables connected
to them. On models with two or more light source ports, visual impairment may occur if light
that is mistakenly emitted from the wrong port enters the eye.
French
AVERTISSEMENT
• Lorsque l’AQ1200x/AQ1205x génère de la lumière, la lumière est émise à travers les
ports de source lumineuse. Ne pas débrancher les câbles de fibre optique connectés. Des
lésions oculaires peuvent être causées si le faisceau lumineux pénètre l’œil.
• Masquer les orifices de source lumineuse sur lesquels aucun câble à fibres optiques n'est
branché. Sur les modèles d'au moins deux orifices de source lumineuse, une déficience
visuelle peut se produire si la lumière
émise par erreur du mauvais orifice pénètre dans l'œil.