43
V2.3.0 OM-UNN valid from 15
th
April 2015
6.5
Daily Maintenance Procedure
6.5.1
Instruction
After every cycle of tests, needles should be cleaned as follows:
A
icon will be shown in RUN page and HOME page after one cycle of tests (See Fig.6-1 and
Fig.6-2). And the boxes in the left side will give red prompts to remind the user to clean needles.
Fig.6-1
Fig.6-2
6.5.2
Daily Cleaning Procedure
Step 1.
Click
icon, a dialog box will pop up to ask the user to insert a service cartridge in position 1
(See Fig.6-3).