Syringe Assembly
Syringe motor moves wrongly.
Disp. Assembly
Dispensing motor moves or dispense wrongly.
Puncher. Assembly
Puncher motor moves or perforates wrongly.
Washer 1 Assembly
Washer motor 1 moves wrongly.
Washer 2 Assembly
Washer motor 2 moves wrongly.
Waste Undefined Error
A generic error has been found at the start of the hydraulic check, it could be
possible damage to the waste circuit or to the result of a defective function of
a sensor. Therefore, it is not possible to continue with the diagnostic
procedure.
Waste Floating Sensor
Error found at the start of the hydraulic check due to damage to the sensor on
the waste container.
Waste Well Cleaning Failed
Error activated:
—when the waste container has exceeded the filling limit and all
functions of the instrument will be disabled.
—from the initial check to test partial emptying of the container.
Waste Pump Fault
Error found at the start of the hydraulic check due to damage to the waste
pump.
Circuit Error
Error found at the start of the hydraulic check due to the disconnection or
damage to the tubing.
Syringe Priming Error
Error found at the start of the hydraulic check due to the system’s inability to
prime the syringe.
Channel 450 Calibration
The dark balancing of optical channel 450 has failed.
Channel 630 Calibration
The dark balancing of optical channel 630 has failed.
Note
Caution
The errors listed in the section below are fatal because the instrument cannot recover the data from the
error(s).Whatever process that is in progress will be aborted. No corrective actions are suggested. Users must
contact Customer Support for field service support.
Identification Errors
The errors described below are detected during the identification procedure of the cartridge, and the
corresponding cartridges with the error will not be processed.