Inputs for sensor calibration
Input signals
Measuring span
Measuring error in
% of the measur-
ing span
1)
Minimum measuring
time
1-channel
measure-
ment
2-channel
measure-
ment
Current signals
0 ... 2 mA
≤ ±0.02 % ±1 digit
-
-
0 ... 5 mA
≤ ±0.02 % ±1 digit
-
-
0 ... 20 mA
≤ ±0.02 % ±1 digit
0.02 s
0.04 s
4 ... 20 mA
≤ ±0.02 % ±1 digit
0.02 s
0.04 s
Voltage signals
DC -18 ... +18 mVDC
≤ ±0.06 % ±1 digit
-
-
DC -35 ... +35 mVDC
≤ ±0.06 % ±1 digit
-
-
DC -75 ... +75 mVDC
≤ ±0.04 % ±1 digit
-
-
DC -150 ... +150 mVDC
≤ ±0.03 % ±1 digit
-
-
DC -300 ... +300 mVDC
≤ ±0.03 % ±1 digit
-
-
DC -500 ... +600 mVDC
≤ ±0.03 % ±1 digit
-
-
DC -500 ... +1,250
mVDC
≤ ±0.03 % ±1 digit
-
-
DC -500 ... +2,500
mVDC
≤ ±0.03 % ±1 digit
-
-
DC -1 ... +5 V
≤ ±0.02 % ±1 digit
-
-
DC -1 ... +10 V
≤ ±0.01 % ±1 digit
0.02 s
0.04 s
Resistance (2-,
3-, or 4-wire)
0 Ω ... 100 Ω
≤ ±0.04 % ±1 digit
-
-
0 Ω ... 1 kΩ
≤ ±0.04 % ±1 digit
-
-
0 Ω ... 10 kΩ
≤ ±0.04 % ±1 digit
-
-
1) The indication of the measuring error applies to ambient temperatures of 20 ... 40 °C and the measuring time
of 1 s.
11. Specifications