50
WIKA operating instructions digital indicator, model DI35-D
EN
140553539.03 03/2020 EN/DE
Input signals
Measuring span
Measuring error
in % of the meas-
uring span
1)
Minimum measuring
time
1-channel
measure-
ment
2-channel
measure-
ment
Voltage signals
DC 0 ... 18 mVDCDC
≤ ±0.06 % ±1 digit
-
-
DC 0 ... 35 mVDC
≤ ±0.06 % ±1 digit
-
-
DC 0 ... 75 mVDC
≤ ±0.04 % ±1 digit
-
-
DC 0 ... 150 mVDC
≤ ±0.03 % ±1 digit
-
-
DC 0 ... 300 mVDC
≤ ±0.03 % ±1 digit
-
-
DC 0 ... 600 mVDC
≤ ±0.03 % ±1 digit
-
-
DC 0 ... 1,250 mVDC
≤ ±0.03 % ±1 digit
-
-
DC 0 ... 2,500 mVDC
≤ ±0.03 % ±1 digit
-
-
DC 0 ... 5 V
≤ ±0.02 % ±1 digit
-
-
DC 0 ... 10 V
≤ ±0.01 % ±1 digit
0.02 s
0.04 s
Thermocouples
Type B, PtRh-PtRh
-100 ... +1,810 °C
≤ ±0.10 % ±1 digit
-
-
Type E, NiCr-CuNi
-260 ... +1,000 °C
≤ ±0.06 % ±1 digit
-
-
Type J, Fe-CuNi
-210 ... +1,200 °C
≤ ±0.05 % ±1 digit
-
-
Type K, NiCr-Ni
-250 ... +1,271 °C
≤ ±0.05 % ±1 digit
-
-
Type L, Fe-CuNi
-200 ... +900 °C
≤ ±0.06 % ±1 digit
-
-
Type N, NiCrSi-NiSI -250 ... +1,300 °C
≤ ±0.06 % ±1 digit
-
-
Type R, PtRh-Pt
0 ... 1,760 °C
≤ ±0.07 % ±1 digit
-
-
Type S, PtRh-Pt
0 ... 1,760 °C
≤ ±0.06 % ±1 digit
-
-
Type T, Cu-CuNi
-240 ... +400 °C
≤ ±0.07 % ±1 digit
-
-
Resistance thermometer
2)
Pt100 (2-/4-wire)
-200 ... +850 °C
≤ ±0.04 % ±1 digit
-
-
Pt100 (3-wire)
-200 ... +850 °C
≤ ±0.04 % ±1 digit
-
-
Pt200 (2-/4-wire)
-200 ... +850 °C
≤ ±0.04 % ±1 digit
-
-
Pt200 (3-wire)
-200 ... +850 °C
≤ ±0.04 % ±1 digit
-
-
Pt500 (2-/4-wire)
-200 ... +850 °C
≤ ±0.04 % ±1 digit
-
-
Pt500 (3-wire)
-200 ... +850 °C
≤ ±0.04 % ±1 digit
-
-
Pt1000 (2-/4-wire)
-200 ... +850 °C
≤ ±0.04 % ±1 digit
-
-
Pt1000 (3-wire)
-200 ... +850 °C
≤ ±0.04 % ±1 digit
-
-
1) The indication of the measuring error applies to ambient temperatures of 20 ... 40 °C and the measuring time
of 1 s.
2)
The indications for Pt100 3-/4-wire apply at a max. lead resistance of 10 Ω.
11. Specifications