32
WIKA operating instructions digital indicator, model DI35-D
EN
7.7 Setting special functions
7.7.1 Function of the digital input
PN
Function
Parameter
53
Function of the digital input
0 = Deactivated
1 = Switching the measuring input
2 = HOLD
3 = TARE
7.7.2 Function of the multi-function key
PN
Function
Parameter
54
Function of the multi-function key
0 = Deactivated
1 = Switching the measuring input
2 = HOLD
3 = TARE
7.7.3 TARE function
The TARE function sets the display value of the stored measuring input to zero. The
difference between zero and the display value is stored as an offset value.
The TARE function overrides a previously programmed offset shift of the
measuring inputs (PN 4/9). Should the TARE function be used during a
measurement, the offset shift has to be reprogrammed.
The TARE function can be activated through the multi-function key or the digital input
(→ see chapter 7.7 “Setting special functions”). The key must be activated for at least 3
seconds.
The tare operation is represented as follows:
7. Description of program numbers
140553539.03 03/2020 EN/DE