Service Manual
Repair Procedures
71
Cuff Fittings
Air system integrity in the NIBP system is essential for accurate measurements. Routine
wear of the O-ring gasket located in the threaded fitting of the cuff hose can result in air
leaks. If you encounter air leaks that cannot be found from the tests in this manual,
replace the O-ring first.
1.
Disconnect the cuff hose from the monitor.
2.
Using pliers, pull off the metal fitting from the cuff hose.
3.
Cut approximately 1/2" off the cuff hose.
4.
Using a small screwdriver, remove the C-clip from the fitting.
5.
Remove the knurled nut from the fitting.
6.
Remove the O-ring gasket and replace it with a new one, being careful not to nick or
otherwise damage the new O-ring.
7.
Reassemble the fitting with the knurled nut and C-clip.
8.
Press the fitting back onto the cuff hose.
WARNING
Use of NIBP channel with leaking cuff hose fittings may result in
inaccurate noninvasive blood pressure measurements.
Cuff
Hose
C-Ring
O-ring
Knurled
Nut
0150-31
Содержание Propaq Encore
Страница 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Страница 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Страница 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 130: ...126 Dynatech Nevada Patient Simulator Modification Welch Allyn Propaq Encore Vital Signs Monitor...