Service Manual
Repair Procedures
51
Replacing the Power Input Fuse
The F1 fuse protects the Recharger board circuitry against excessive current at the dc
input connector. You do not need to disassemble the Propaq Encore to replace this fuse.
Check fuse F1 if the ac power adapter is functioning properly and the following two
conditions exist:
•
the green LED charging indicator on the monitor’s right side panel does not
illuminate
•
the monitor battery does not charge
Recharger
Board
Right
Side Panel
Left
Side Panel
Rear
Chassis
0696-12
DEFIB SYNCHRO
MONITOR
EKG x 1000
!
3A
2AG
12-28V, 3A
!
!
F1