Service Manual
Functional Verification
17
Sequence for All Other Software Versions
1.
Press ECG/RESP > MORE. Then press CHANGE to select the four HR/PR TONE
levels. Verify that the tone volume changes and goes off when turned OFF. Set the
volume to LOW.
2.
Press MAIN MENU.
3.
Software versions 2.5X:
Press SETUP > MORE > MORE > SERVICE > YES > MORE
> MORE > SETTINGS > ALL ALRM. Verify alarm status displays "TURNING OFF ALL
AUDIBLE VITAL SIGNS ALARMS", "ARE YOU SURE?", YES or NO. Then press MAIN
MENU.
4.
All software versions:
Press SETUP > MORE > MORE (> TIME/DAY for monitor
software less than version 2.0) and check that the displayed time and date are correct.
If incorrect, press NEXT and UP or DOWN to select and change the time and date,
and then set by pressing the ENTER button.
5.
Turn off the monitor.
ECG/RESP Channel and Alarm Indicators
The following procedures verify the ECG channel and the alarm indicator drivers. The
setup procedure below is used for both the visual and the electrical checks that follow.
Setup
1.
Turn on the monitor and verify that it powers up correctly.
2.
Put the monitor into FACTORY ADULT MODE.
Software version 2.XX:
SETUP > MORE > CHANGE > SETUP > USE NOW > YES
Software version 1.XX:
SETUP > MORE > CHANGE > ADULT > YES > MORE >
PROGRAM > DEFAULT > YES
3.
Press MAIN MENU.
Press
Result
FREEZE/UNFREEZE
Freezes the waveforms
FREEZE/UNFREEZE
Unfreezes the waveforms
SETUP
Changes the menu
ALARMS
Changes the display to ALARM STATUS
SUSPEND (on menu)
SUSPEND changes to RESUME
ALARM SUSPEND/RESUME (button)
Changes RESUME to SUSPEND
ALL ALRM
Changes display: Turn All Off, Are You Sure, Y or N
LIMITS
Changes display to HR Limits
NEXT
Moves cursor block: Next block within selected parameter limits, NEXT
PAGE to next parameter
MAIN MENU (button)
Returns to the main menu
NIBP START/STOP (button)
Starts the NIBP pump
NIBP START/STOP (button)
Stops the NIBP pump
Содержание Propaq Encore
Страница 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Страница 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Страница 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 130: ...126 Dynatech Nevada Patient Simulator Modification Welch Allyn Propaq Encore Vital Signs Monitor...