Service Manual
Repair Procedures
57
3.
Spread the casing open to an angle of about 80 degrees.
4.
Detach the following cables from the front chassis (indicated in figure below):
•
Cable to P5 on the Analog board, from the left side panel ECG connector on the
rear chassis.
•
Cable to P6 on the Analog board, from the left side panel IPB connector on the
rear chassis.
•
Cable to P7 on the Analog board, from the left side panel temperature jack on the
rear chassis.
•
Cable to P8 on the lower (Digital) board, from the right side panel on the rear
chassis.
•
Cable to P9 on the Digital board, from the right side panel on the rear chassis.
DISCONNECT 2
UNDERLYING
DIGITAL BOARD
DISCONNECT
3 CABLES
CABLES FROM