Service Manual
Repair Procedures
65
6.
To carefully remove the display:
a.
Using a flat blade screwdriver, each person should hold the silicone gasket down
where it is visible between the posts at the top and bottom of the chassis (see
figure above).
b.
With their free hand, each person should gently lift one side of the display
module, attempting to keep the silicone gasket in place on the display filter.
If the gasket lifts off the display filter at any time, inspect the filter beneath the section
where the gasket lifted to ensure the conductive coating around the perimeter of the filter
is still intact. Try not to disturb the rest of the gasket/filter interface as you do this. To
inspect the filter, look for a slight color variation between the perimeter conductive
coating and the rest of the filter (see the figure below). (Tilting the front assembly at
different angles under a bright light can help you see the color variation of the conductive
coating.) The slight color variation should be continuous around the perimeter under the
gasket. If there is a break in this coating, you must replace the front assembly with the
display filter.
Caution
In the next step, if the gasket lifts off the display filter at any point, you
must inspect the display filter to verify that the conductive coating on the filter is
still intact. See the steps below.
HOLD THE GASKET DOWN
AT THESE POINTS ON BOTH
SIDES OF THE BOARD.
REMOVE 4 SCREWS
AT CORNERS
Содержание Propaq Encore
Страница 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Страница 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Страница 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 130: ...126 Dynatech Nevada Patient Simulator Modification Welch Allyn Propaq Encore Vital Signs Monitor...